4 Notes – Maintenance – Faults
750 Autosampler
48
4.4.3
Remote control error codes
If the remote control for the 750 Autosampler is switched on via the
RS232 interface then each error will produce a corresponding error
code. For each of the errors listed below information is provided as to
whether the error can be remedied (R) or if it is a fatal system error (F).
The Autosampler remains ready for operation with R-errors and the op-
erator can enter a new command. With F-errors it is no longer possible
to use the Autosampler. The instrument must be switched off and re-
paired (see chap. 4.4.2).
Code Error
Type
0
No error, autosampler is not busy and is ready for a
command
.
–
-1
Invalid auxiliary output contact number request
.
R
-2
Invalid auxiliary output command request
.
R
-4
Request invalid because autosampler is
busy performing an operation
.
R
-5
Invalid buzzer request
.
R
-6
Invalid command request. Autosampler did
not recognize command
.
R
-8
Invalid display request
.
R
-12
Present needle position is unknown or invalid
.
R
-13
Requested needle position is invalid
.
R
-21
Invalid vial number request
.
R
-22
Transfer needle is not at a valid vial position, so it
cannot be lowered
.
R
-27
Invalid injection valve position request
.
R
-28
Motor drive error on x drive (left-right).
F
-29
Motor drive error on y drive (forward-back).
F
-30
Unable to determine vial position or vial position is
invalid
.
R/F
-31
Motor drive error on z drive (up-down).
F
-32
Motor stall error on z drive
.
F
-35
Error, origin unknown
.
F
-36
Autosampler stopped locally
.
R
-43
Step position requested by MVZ not valid
.
–
-47
Step position requested by MVX or MVY not valid
.
–
-48
Not a valid X/Y step position
.
–
-49
Serial communications error
.
–