8 remote control – Metrohm 750 Autosampler User Manual
Page 39

3 Operation
750 Autosampler
36
3.8
Remote control
REM/LOC
6
By pressing the
menu for switching between normal operation (Local) and remote-
controlled operation via the RS232 interface (Remote) appears. The key
opens the following menu:
REMOTE/LOCAL
Local
RS-232
IEEE-488
Exit
Remote control/Normal operation
Normal operation
Remote control via RS232
Remote control via IEEE-488
Return to standby mode
The individual menu points are selected by pressing the <
б
б
> and
<
в
в
> keys. The individual inquiries of a menu point are accessed using
the
log items which appear under
displays are the default values, the possible entry values or ranges are
shown below the display.
REMOTE/LOCAL
à
Local
Switch to normal operation
After confirmation of this command with
is switched from operating under remote control via RS232 to
normal operation.
REMOTE/LOCAL
à
RS-232
Switch to remote control via RS232
After confirmation of this command with
display appears:
Select Baud Rate
<4800>
300,600,1200,2400,
4800,9600,19200,
38400
Entry of baud rate
Entry of baud rate in bits/s for the RS232 interface. The value
entered here must also be set on the external controller. After
the baud rate has been confirmed with
lowing display appears:
Select Parity
Even,Odd,None
Entry of parity
Entry of parity for the RS232 interface. The value entered
here must also be set on the external controller. After the
parity has been confirmed with
display appears: