3 Operation

750 Autosampler

32

EDIT METHOD # N

à

Delete Step

Delete program step
Allows program steps to be deleted. After

is pressed the following display appears:

DELETE A STEP

Enter Step Number
#N

Select program step
Entry of the number of the program step which is

to be deleted. After is pressed the

following display appears:

Delete Step # N?

to Delete
to Abort

Confirmation of deletion process

Delete selected program step N.

Abort deletion process.

METHOD MENU

à

Run Method

Running a method
Allows a method to be started. After is pressed the

following display appears:

Enter Method Number

# N

1

10

Select method
Entry of the method number N which is to be started. After

confirmation with the selected method is carried

out exactly as if it has been started with the

key (details see chap. 3.6).

METHOD MENU

à

Copy Method

Copy a method
Allows a method to be copied. After is pressed the

following display appears:

COPY METHOD

Copy FROM Method
# N

1

10

Source method
Entry of the method number N which is to be copied. After

confirmation with the following display appears:

COPY METHOD

Copy TO Method
# M

1

10

Target method
Entry of the method number M which is to be overwritten

with the selected source method. After confirmation with

the method will be copied provided that the

target method is empty.
If the target method already contains program steps then the

following safety inquiry will appear: