Metrohm 864 Robotic Balance Sample Processor User Manual
Page 56
![background image](/manuals/586625/56/background.png)
3.17 Adjusting the rack and the robotic arm
■■■■■■■■■■■■■■■■■■■■■■
48
■■■■■■■■
864 Robotic Balance Sample Processor
The last position on the sample rack is conceived as Adjusting posi-
tion.
3 Define the special beaker on the 6.2068.010 sample rack
■
Select the lowest line (Special beaker 16) and click [Edit].
■
Under Rack position, select 61.
■
Close the dialog window with [OK].
■
Close the rack data table with [OK].
Moving to the positioning reticle
Proceed as follows:
1 Open the manual control
■
In the Manual control in tiamo
™, click the tab Move.
2 Move to special beaker 16
■
Under Rack position, select the target position Special beaker
16 and click [Start].
It is also possible to specify the target position as absolute rack posi-
tion. For the 6.2068.010 sample rack, the adjusting position is the
no. 61.
3 Move to the adjusting position
■
Unscrew the FEP tubing from the adapter on the robotic arm.
■
Insert the PEEK capillary supplied from above into the adapter.
The capillary must protrude below from the adapter by around
0.5 cm (see following illustration)
■
Under lift position enter 100 mm as target position and click
[Start].