Figure 26, Installing the ionizing bar – Metrohm 864 Robotic Balance Sample Processor User Manual
Page 49
■■■■■■■■■■■■■■■■■■■■■■
3 Installation
864 Robotic Balance Sample Processor
■■■■■■■■
41
■
Under COM Port, select the connector on the PC the connection
cable was connected to.
■
Enter the values listed above as RS-232 parameters.
■
Click [Connect].
The dialog window Establish connection is displayed.
■
Press the [Print] key on the balance.
The data sent by the balance is displayed.
■
Close the dialog window with [OK].
The input fields in the properties window are now locked. This
means that the connection between balance and PC has been
established.
■
Close the properties window with [OK].
Installing the ionizing bar
In order to eliminate electrostatic loads while weighing, an ionizing bar
must be installed as close to the balance as possible. Proceed as follows:
6.2057.160
1
2
3
Figure 26
Installing the ionizing bar
1 Fasten the holder
■
Fasten the 6.2057.160 holder on the assembly rail with the aid of
a hexagon key.
The position must be selected in such a way that the recess of the
holder surrounds the weighing column.
2 Insert the screws
■
Insert the two accompanying screws from above into the holding
rail of the ionizing bar.