2 connecting the swing head, 3 mounting the deflector and collection container, Connecting the swing head – Metrohm 864 Robotic Balance Sample Processor User Manual
Page 23: Mounting the deflector and collection container, Figure 6, Connecting the mains cable, Figure 7, Connecting swing head
![background image](/manuals/586625/23/background.png)
■■■■■■■■■■■■■■■■■■■■■■
3 Installation
864 Robotic Balance Sample Processor
■■■■■■■■
15
P: 115W U: 100 - 240 V f: 50 - 60 Hz
Nr
.
Figure 6
Connecting the mains cable
3.2.2
Connecting the Swing Head
Take care to ensure that the Swing Head is connected before the instru-
ment is started up. Check the connection cable.
The connection socket (Mini DIN) for the Swing Head drive is each located
on the rear of the tower next to the stirrer connector.
Figure 7
Connecting Swing Head
If the Swing Head is not connected, connect it as follows:
1 Plug in the cable
■
Guide the connection cable of the Swing Head through the guide
chain of the tower (see Chapter 3.3, page 17).
■
Plug the Mini DIN plug into the socket 'Swing Head'.
3.2.3
Mounting the deflector and collection container
When using pipetting tips, injection needles and disposable filters, these
need to be stripped off the robotic arm again afterwards. For this pur-
pose, a deflector and a collection container are mounted on the fastening
plate on the right-hand side of the tower 1. Proceed as follows:
1 Mount the deflector
■
Screw the 6.2058.070 deflector tightly to the fastening plate with
the screws and washers supplied.
We recommend selecting the highest position on the fastening
plate. For positioning, the deflector can be shifted laterally as
required.