Metrohm 864 Robotic Balance Sample Processor User Manual
Page 45
■■■■■■■■■■■■■■■■■■■■■■
3 Installation
864 Robotic Balance Sample Processor
■■■■■■■■
37
3 Prepare the titrant
■
Fasten a filling tubing and an adsorber tube (filled with wadding)
on a 6.3032.220 20 mL dosing unit.
■
Screw a dosing unit on a bottle with titrant.
A 6.1608.030 clear glass bottle (1 L, with GL45 thread) can be
used.
■
On Port 1 of the dosing unit, fasten one of the 6.1805.130 FEP
tubings (100 cm), which are already mounted on the titration
head, see previous chapter.
4 Mount the drive for the titrant
■
Attach an 800 Dosino drive to the dosing unit.
■
Connect the connection cable to the connector MSB 1 of the
Titrando.
5 Prepare the solvent
■
Fasten a filling tubing and an adsorber tube (filled with wadding)
on a 6.3032.250 50 mL dosing unit.
■
Screw a dosing unit onto the 6.1608.070 clear glass bottle (2 L,
with GL45 thread) filled with solvent.
■
On Port 1 of the dosing unit, fasten one of the 6.1805.130 FEP
tubings (100 cm), which are already mounted on the titration
head, see previous chapter.
6 Mount the drive for the solvent
■
Attach an 800 Dosino drive to the dosing unit.
■
Connect the connection cable to the connector MSB 2 of the
864 Robotic Balance Sample Processor.
7 Connecting the electrode cable
■
Connect the 6.2104.030 electrode cable which is already con-
nected to the electrode to the connector Ind. on the rear of the
Titrando.
8 Connect the Titrando
■
Connect a 6.2151.000 connection cable to the Titrando (Con-
troller connector).
■
Connect the cable to a USB connector on the Sample Processor
or on the PC.
After that, the Titrando is recognized by the tiamo
™ software and
can be added to the device table. The same applies for the dosing