2 cable connections, Cable connections, Figure 28 – Metrohm 864 Robotic Balance Sample Processor User Manual
Page 51: Cabling
■■■■■■■■■■■■■■■■■■■■■■
3 Installation
864 Robotic Balance Sample Processor
■■■■■■■■
43
5
Pipetting tubing
6.1805.130 FEP tubing
800 Dosino (on empty bottle P)
— 864
tower 1, pipetting tip adapter
6
Titration tubing
6.1805.120 FEP tubing
800 Dosino (titrant T)
— 864 tower 2, titra-
tion tip
7
Solvent feed line (Solvent)
6.1805.120 FEP tubing
800 Dosino (solvent S)
— 864 tower 2, dos-
ing tip
3.14.2
Cable connections
1
2
5
6
9
8
10
11
12
843
Ion
B
864
809
PC
tiamo
TM
800
800
T1
T2
USB
COM1
3 4
7
S
T
P
Figure 28
Cabling
1
Pump connection
6.2141.300 connection cable.
843 Pump Station (Remote 2)
— 864:
tower 2 (pump 1 and pump 2)
2
Ionizer connection
Ionizer (control device)
— Ionizing bar
The connection cable is supplied with the
ionizer.
3
Stirrer cable
Connection cable of the rod stirrer (802 Stir-
rer)
Connector on tower 2 (Stirrer)
4
Swing Head cable
Connection cable of the 786 Swing Head
Connector on tower 2 (Swing Head).
5
Balance connection
RS-232 serial connection cable. Supplied
with the balance.
Balance (RS 232)
— PC (COM1)
6
USB connection
6.2151.000 controller cable (USB A-MiniDIN
8p)
864 (Controller)
— PC (USB)