12 setting up the dosing devices and the titrator, Figure 24, Setting up the dosing devices and the titrator – Metrohm 864 Robotic Balance Sample Processor User Manual
Page 44
3.12 Setting up the dosing devices and the titrator
■■■■■■■■■■■■■■■■■■■■■■
36
■■■■■■■■
864 Robotic Balance Sample Processor
3.12
Setting up the dosing devices and the titrator
Supplied with the 864 Robotic Balance Sample Processor are three dosing
drives and a titrator. They can be set up in a compact arrangement
according to the following illustration.
1
2
3
4
5
6
Figure 24
Setting up the dosing devices and the titrator
Setting up the dosing units and the Dosinos
Proceed as follows:
1 Install the dosing unit for pipetting
■
Screw a 6.3032.220 20 mL dosing unit onto an empty
6.1608.030 clear glass bottle (1 L, with GL45 thread).
■
On Port 1, fasten the 6.1805.130 FEP tubing (120 cm) which
is connected to the pipetting tip adapter on tower 1, (see
"Mounting the adapter", page 26).
2 Mount the pipetting drive
■
Attach an 800 Dosino drive to the dosing unit.
■
Connect the connection cable to the connector MSB 1 of the
Sample Processor.
This is the pipetting drive. The Dosino pumps air in order to aspirate
the sample into the pipetting tip and then to expel this back out
again.