beautypg.com

11 connecting the tower stirrer, Figure 21, Rod stirrer 802 stirrer – Metrohm 864 Robotic Balance Sample Processor User Manual

Page 43: Figure 22, Magnetic stirrer 741 stirrer, Figure 23, Connecting the tower stirrer

background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

864 Robotic Balance Sample Processor

■■■■■■■■

35

4 Connect the dosing tubings

Connect the two 6.1805.120 dosing tubings of the titrant and
solvent to the pre-mounted dosing tips on the titration head.

3.11

Connecting the tower stirrer

A DIN socket for connecting a rod stirrer (802 Stirrer) or a magnetic stir-
rer (741 Stirrer) is located on the rear side of the tower.

Figure 21

Rod stirrer 802 Stirrer

Figure 22

Magnetic stirrer 741 Stirrer

Take care to observe correct orientation of the contact pins when plug-
ging in the stirrer connection cable. The rib on the outside of the plug
must match the reference mark (on the left) on the socket.

Figure 23

Connecting the tower stirrer

Note

If an MSB stirrer is connected to the MSB1 or MSB2 socket, then the
stirrer connector on tower 1 or tower 2 cannot be used, because the
tower stirrers are internally controlled as well via MSB1 or MSB2.