3.2 Carrying out titrations

■■■■■■■■■■■■■■■■■■■■■■

32

■■■■■■■■

870 KF Titrino plus

Select the menu item Calculation and open the corresponding
dialog with [OK].

The calculation formula is defined for every method template and
cannot be modified. Depending on the unit of the sample size
and the result, only the conversion factors FCT and DIV need to
be adjusted.

Select the line with the calculation formula and press [OK].

This conversion table is for purposes of information only. It is not
possible to apply values taken from there.

Find the desired result unit and the unit of the sample size in the
table and make a note to yourself of the factor and divisor con-
tained in this line.

Close the table with [BACK].

Enter the values for Factor (FCT) and Divisor (DIV) according to
the conversion table shown above.

4 Activate the statistics

With [BACK], switch to the parameters dialog.

Select the menu item Statistics and open the corresponding dia-
log with [OK].

Select the parameter Statistics and confirm with [OK].

Select on and confirm with [OK].
The parameter Number of determinations is additionally dis-
played.

Select the parameter Number of determinations and open the
editing dialog with [OK].

Enter the number of determinations to be carried out, in this
example 3.

Apply the value with Accept or [BACK].