2.5 Setting up tubings and tips

■■■■■■■■■■■■■■■■■■■■■■

14

■■■■■■■■

870 KF Titrino plus

Attaching the exchange unit

1 Slide the exchange unit from the front onto the 870 KF Titrino plus

and push all the way to the rear.

It must snap in audibly.

Inserting the titration tip

1

1 Insert the titration tip of the exchange unit into the remaining open-

ing at the rear on the vessel lid.

The end of the tip should be located just above the stirring bar, but is
not permitted to touch it.