3.2 Carrying out titrations

■■■■■■■■■■■■■■■■■■■■■■

28

■■■■■■■■

870 KF Titrino plus

2 Inject water standard

Note

Calculate the amount of the water standard in such a way that a
titration consumption of 10…90 % of the cylinder volume will
result.

Press [START].
Conditioning is stopped. The request for adding the water stand-
ard will be displayed for 8 s.
The water standard should be injected during this time.

Inject the water standard.

Afterward the request for the sample size appears:

3 Enter the sample size

Open the editing dialog with [OK].

Enter the sample size and apply with Accept or [BACK].

4 Start the titration

Press [START].