Setting up tubings and tips

Setting up the aspiration and dosing tip

1

2

3

4

5

6.2730.030

6.1543.110

6.1543.120

1 Place the three stoppers 6.2730.030 (including O-rings, but without

nipples) in the rear openings of the vessel lid.

2 Screw on the dosing tip 6.1543.110 to the M8 PTFE tubing of the

solvent bottle (amber glass).

3 Insert the dosing tip through the stopper in the middle rear opening

of the vessel lid.

4 Screw the aspiration tip 6.1543.120 onto the M8 PTFE tubing of the

aspiration bottle (clear glass).

5 Insert the aspiration tip through the stopper in the right rear opening

of the vessel lid.

The end of the tip must touch the bottom of the vessel, but it must
not inhibit the action of the stirring bar.

As an alternative to the aspiration tip 6.1543.120, the aspiration tip
6.1821.070 can also be used with an M10 screw nipple.