Page 30

background image

3.2 Carrying out titrations

■■■■■■■■■■■■■■■■■■■■■■

24

■■■■■■■■

870 KF Titrino plus

Select the parameter Solution and confirm with [OK].
The list of the configured solutions is displayed:

Select the newly configured solution and apply with [OK].

3 Adjust the calculation

With [BACK], switch to the parameters dialog.

Select the menu item Calculation and open the corresponding
dialog with [OK].

The calculation formula is defined for every method template and
cannot be modified.

Select the parameter Factor (FCT) and open the editing dialog
with [OK].

Enter the water content of the water standard used in mg/g (see
certificate of the water standard).

Apply the value with Accept or [BACK].

Note

In the manual you will find details as to how the conversion factor
FCT is to be adjusted if you do not use a water standard to carry
out the titer determination.

4 Activate the statistics

With [BACK], switch to the parameters dialog.