2.1 Setting up and connecting the instrument

■■■■■■■■■■■■■■■■■■■■■■

4

■■■■■■■■

870 KF Titrino plus

Connecting the titration stand to the MSB 1

Observe the marking on the socket.

Connecting the electrode cable

6.2104.020

The electrode cable 6.2104.020 is protected against accidental disconnec-
tion of the cable by means of a "pull-out protection" feature. If you wish
to remove the plug, then you must first retract the outer plug sleeve.

Connecting a printer or other USB instruments

Adapter 6.2151.100 is supplied for the purpose of connecting instruments
or storage media or similar that are equipped with a USB interface.

6.2151.100

A printer can be connected directly to the adapter (with a connecting
cable 6.2151.020), as can a USB keyboard or a USB storage medium (USB
Memory Stick for storing methods, etc.).

If several instruments are to be connected, then it is recommended that a
USB hub (available from speciality dealer outlets) be used and that this be
connected to the adapter 6.2151.100. The USB hub then functions as a
distributor to which several USB instruments, etc. are connected.

The cable 6.2151.120 is available as a possible alternative; using this
cable, a printer, a USB hub or an RS 232/USB Box (see next section) can be