The titration starts and the curve is displayed:

After the completion of the titration, the results dialog is displayed. If
a printer is connected, then the report will now be printed out. Con-
ditioning is restarted automatically.

5 Carry out the remaining determinations

Carry out the remaining titer determinations in the same way.

6 Show the statistics data

After the final determination, select the function Statistics in the
function bar and confirm with [OK].
The statistical overview is shown.

The mean value (Mean( )), the absolute and the relative standard
deviation (s abs and s rel) are displayed in the overview. For the
mean value, the number of individual results from which it has
been calculated is displayed. In this example it is 5. Statistics
shows how many determinations have already been carried out
and how many determinations are to be carried out in total. All
five determinations were carried out in this example.

In the function bar, select the Details function and confirm with
[OK].