2.2 Setting up the KF titration cell

■■■■■■■■■■■■■■■■■■■■■■

8

■■■■■■■■

870 KF Titrino plus

Equipping the KF titration cell

6.2730.020

6.2730.010

6.0338.100

6.1403.040

6.2104.020

1

2

3

4

5

6

The KF titration cell can now be equipped as follows:

1 Introduce the septum stopper 6.2730.020 (with septum inserted)

into the front opening of the vessel lid.

2 Insert the O-rings of the electrode and of the drying tube into the

middle openings of the vessel lid.

3 Screw the two screw nipples 6.2730.010 into the openings with the

O-rings. Do not screw them in too tightly.

4 Introduce the double Pt-wire electrode 6.0338.100 into the left-hand

opening and then tighten the screw nipple until it seals.

5 Screw the electrode cable 6.2104.020 tightly onto the electrode.

6 Introduce the filled drying tube 6.1403.040 on the right of the elec-

trode into the remaining opening and then tighten the screw nipple
until it seals.