4.2 Remote control commands

756/831 KF Coulometer, Instructions of Use

87

Info.ActualInfo.Titrator.I read

only

Info.ActualInfo.Titrator.Pot read

only

Info.ActualInfo.Titrator.IPulse read

only

$Q sends the current values.

.CyclNo: Cycle number of the voltage measurement cycle, e.g. "127". From the

cycle number and the cycle time (see 4.2.2.77), a time frame can be set

up. The cycle number is set to 0 on switching on the instrument and on

every start. It is incremented as long as the instrument remains switched

on.

.Water:

Total water in ug, e.g. "62.313"

.Meas:

Measured value in mV at the indicator electrode, e.g. "104.2".

.dWaterdt: Rate or drift in ug/min, e.g. "23.0".

.I:

Total charge in mA·s, e.g. "667.48".

.Pot:

Voltage at generator electrode.

0 means "undefined", 1 means <14 V, 2 means 14...28 V, 3 means

>28 V.

.IPulse:

Current of actual pulse.

1 means 100 mA, 2 means 200 mA, 3 means 400 mA.

OV will be sent for "overrange".

4.2.2.72. Info.ActualInfo.MeasPt.Index read

only

Info.ActualInfo.MeasPt.X read

only

Info.ActualInfo.MeasPt.Y read

only

Info.ActualInfo.MeasPt.Z1 read

only

Info.ActualInfo.MeasPt.Z2 read

only

Info.ActualInfo.EP.Index read

only

Info.ActualInfo.EP.X read

only

Info.ActualInfo.EP.Y read

only

$Q sends the last entry into the measuring point list (.MeasPt) or the last entry into

the list of EP.

.MeasPt.X Time in s, e.g. "14".

.MeasPt.Y Water in ug, e.g. "27.5".

.MeasPt.Z1 Measured value in mV, e.g. "160.3".

.MeasPt.Z2 Rate in ug/min, e.g. "100.5".

.EP.X

Water in ug, e.g. "26.6".

.EP.Y

Measured value in mV, e.g. "98.6".

4.2.2.73. Info.ActualInfo.Oven.HeatTime read

only

Info.ActualInfo.Oven.SampleTemp read

only

Info.ActualInfo.Oven.LowTemp read

only

Info.ActualInfo.Oven.HighTemp read

only

Info.ActualInfo.Oven.GasFlow read

only

Info.ActualInfo.Oven.UnitFlow read

only

$Q sends the current values from a connected KF Oven. If no Oven is connected, the

values are empty.

.HeatTime: Heating time of sample in s.

.SampleTemp: Nominal sample temperature in °C.

.LowTemp: Lowest temperature during the sample heating time in °C.

.HighTemp: Highest temperature during the sample heating time in °C.

.GasFlow: Average gas flow during sample heating time.