4.2 Remote control commands
756/831 KF Coulometer, Instructions of Use
75
.Pause:
Time in s. During this time, there is no generation of current.
.ExtrT:
Extraction time in s. During this time controlling occurs but the titration
will not be stopped.
4.2.2.8. Mode.Parameter.TitrPara.StartDrift
1...20..999
StartDrift in ug/min. Drift for "conditioning ok" and start of titration possible.
4.2.2.9. Mode.Parameter.TitrPara.Ipol 2,
5,
10, 20, 30
Mode.Parameter.TitrPara.PolElectrTest
ON, OFF
.Ipol:
Selection of polarization current.
If the test for polarized electrodes is switched on, it is performed on changeover from
the inactive state to an active state (titration or conditioning).
4.2.2.10. Mode.Parameter.TitrPara.Temp -170.0...25.0...500.0
Titration temperature in °C.
4.2.2.11. Mode.Parameter.TitrPara.TDelta
1...2...999 999
Time interval in s for the entry of a measurement point in the list of measured points.
4.2.2.12. Mode.Parameter.TitrPara.TMax
1...999 999, OFF
Maximum titration time in s. After this time, the titration will be stopped.
4.2.2.13. Mode.Parameter.Statistics.Status ON, OFF
Mode.Parameter.Statistics.MeanN
2...20
Mode.Parameter.Statistics.ResTab.Selected original, delete n,
delete
all
Mode.Parameter.Statistics.ResTab.DelN
1...20
Entries for the statistics calculations.
.Status:
On/off switching. Requirement for statistics calculations is a valid as-
signment, see 4.2.2.26.
.MeanN: Number of individual results for statistics calculations.
.ResTab.Select: Selection of the table for the statistics calculations.
original: Original table. The original table is (again) set up, i.e. any indi-
vidual results which have been deleted are reincorporated in the sta-
tistics calculations.
delete n: Single result lines are removed from the statistics calculation.
All results of the corresponding line in the statistics table are de-
leted. Specification of the line number in .ResTab.DelN.
delete all: Clear entire statistics table. The results can not be reactivated.
.ResTab.DelN: Specification of the line number to be deleted.
4.2.2.14. Mode.Parameter.Presel.Cond
ON, OFF
Mode.Parameter.Presel.DCor.Type
auto, man., OFF
Mode.Parameter.Presel.DCor.Value 0.0...99.9
.Cond:
Conditioning ON/OFF