Metrohm 756 KF Coulometer User Manual
Page 82
4.2 Remote control commands
756/831 KF Coulometer, Instructions for Use
74
4.2.2 Description of the remote control commands
4.2.2.1. Mode $G,
$S
Start and stop ($G, $S) of the current method (4.2.2.3)
$G also serves to continue after inquiries of identifications and sample size after the
start (see 4.2.2.15)
4.2.2.2. Mode.Select KFC, KFC-B, BLANK, GLP
Selection of the standard mode.
If a method is selected from the method memory, the node &Mode.Select is over-
written with the mode of the corresponding user method.
4.2.2.3. Mode.Name read
only
Name of the current method in the working memory. $Q sends 8 ASCII characters.
Standard methods carry the name ********. The node can be set read + write,
see 4.2.2.66.
4.2.2.4. Mode.Parameter.CtrlPara.EP
0...50...
±
2000
Setting of the EP in mV.
4.2.2.5. Mode.Parameter.CtrlPara.Control content,
special
Mode.Parameter.CtrlPara.Special.Dyn
1...70...2000
Mode.Parameter.CtrlPara.Special.MaxRate 1.5...2240,
max.
Mode.Parameter.CtrlPara.Special.MinRate 0.3...15...999.9, min.
Mode.Parameter.CtrlPara.Special.Stop.Type drift,
rel.drift
Mode.Parameter.CtrlPara.Special.Stop.Drift 1...5...999
Mode.Parameter.CtrlPara.Special.Stop.Drift 0...5...999
Parameters for setting "special" (4.2.2.5):
.Dyn:
Dynamics in mV.
.MaxRate: Maximum allowed titration rate in ug/min. Max. means maximum possi-
ble rate.
.MinRate: Minimum titration rate in ug/min.
.Type:
Type of stop criterion after drift or switch-off delay time.
.Drift:
Stop drift in ug/min. Applies when "drift" has been selected.
.RelDrift: Relative stop drift in ug/min. Applies when "rel.drift" has been selected.
Stops if the drift reaches the current drift at the start of the method plus
the rel.drift value.
4.2.2.6. Mode.Parameter.TitrPara.Direction +,
-, auto
Titration direction.
"auto" means the titration direction is determined automatically by the instrument.
4.2.2.7. Mode.Parameter.TitrPara.Pause
0...999 999
Mode.Parameter.TitrPara.ExtrT
0...999 999