Metrohm 756 KF Coulometer User Manual
Page 84

4.2 Remote control commands
756/831 KF Coulometer, Instructions for Use
76
.DCor.Type: Type of drift take-over for the drift correction. auto: Take-over of the drift
value at start.
.DCor.Value: Drift value for the manual drift correction.
4.2.2.15. Mode.Parameter.Presel.IReq
id1, id1&2, all, OFF
Mode.Parameter.Presel.SReq
value, unit, all, OFF
Mode.Parameter.Presel.ReqTitr
ON, OFF
Automatic inquiries after the start of the determination. From such an inquiry, the de-
termination continues if the requested entry/entries is/are made, e.g.
&SmplData.OFFSilo.Id1 (see 4.2.2.56) or with &M $G, see 4.2.2.1.
.ReqTitr: Current generation starts during requests (with ON).
4.2.2.16. Mode.Parameter.Presel.SampleUnit
g, ...up to 5 ASCII
Method specific sample unit, i.e. when the method is loaded, the current unit of the
sample size is overwritten by the unit from the method.
4.2.2.17. Mode.Parameter.Presel.LimSmplSize.Status ON,
OFF
Mode.Parameter.Presel.LimSmplSize.LoLim
0.0...999 999
Mode.Parameter.Presel.LimSmplSize.UpLim 0.0...999 999
Limit control for the sample size.
4.2.2.18. Mode.Parameter.Presel.Id1Text
id1/C21, 10 ASCII characters
Mode.Parameter.Presel.Id2Text
id2/C22, 10 ASCII characters
Mode.Parameter.Presel.Id3Text
id3/C23, 10 ASCII characters
Text for sample identifications.
4.2.2.19. Mode.Parameter.Presel.Cell
no diaph., diaphragm
Mode.Parameter.Presel.GenI 100,
200,
400, auto
.Cell:
Type of generator electrode.
.GenI:
Current at the generator electrode in mA. "auto" means that the current is
switched in the course of determinations.
Default: 400 mA for cells without diaphragm, auto for cell with dia-
phragm.
4.2.2.20. Mode.Parameter.Presel.Oven COM1,
COM2,
no
If an Oven is connected, its results will be incorporated into the result report of the
Coulometer.
If there is no Oven connected via RS232, this parameter has to be on "no".
4.2.2.21. Mode.Parameter.Presel.ActPuls
first, all, cond., OFF
Output of a pulse on the I/O line "Activate", see page 132.
4.2.2.22. Mode.Def.Formulas.1.Formula
H2O, CXX, RSX, +, -, *, /, (, )
Mode.Def.Formulas.1.TextRS
8 ASCII characters