beautypg.com

Metrohm 756 KF Coulometer User Manual

Page 100

background image

4.2 Remote control commands

756/831 KF Coulometer, Instructions for Use

92

Setup.SendMeas.Assembly.Pot ON, OFF

Setup.SendMeas.Assembly.IPulse ON, OFF

Setup.SendMeas.Assembly.Bur.V ON, OFF

Selection of the values from Assembly for the output in the set time interval (see

4.2.2.92):

.CyclNo: Cycle number of the potential measurement. Together with the cycle

time (4.2.2.77), a time frame can be set up.

The cycle number is set to 0 on switching on the instrument and it is al-

ways incremented as long as the instrument remains switched on.

.I:

Total charge in mA·s associated to the cycle number, e.g. "667.48".

.Meas:

Measured value in mV associated to the cycle number, e.g. "104.2".

.Pot:

Voltage at generator electrode associated to the cycle number.

0 means "undefined", 1 means <14 V, 2 means 14...28 V, 3 means

>28 V.

.IPulse:

Current of pulse associated to the cycle number.

1 means 100 mA, 2 means 200 mA, 3 means 400 mA.

.Bur.V:

Dosed volume of connected Dosino in ml, e.g. "5.234".

The unit "assembly" must be preset (see 4.2.2.92).

4.2.2.94. Setup.SendMeas.Titrator.CyclNo ON, OFF

Setup.SendMeas.Titrator.Water ON, OFF

Setup.SendMeas.Titrator.Meas ON, OFF

Setup.SendMeas.Titrator.dWaterdt ON, OFF

Setup.SendMeas.Titrator.I ON, OFF

Setup.SendMeas.Titrator.Pot ON, OFF

Setup.SendMeas.Titrator.IPulse ON, OFF

Selection of the values from the titrator which are sent in the set time interval (see

4.2.2.91):

.CyclNo: Cycle number. Together with the cycle time (4.2.2.78), a time frame can

be set up. The other data belong to the corresponding cycle number. The

cycle number is set to 0 at the start of a method and it is incremented

until the end of the method.

.Water:

Total water associated to the cycle number in ug, e.g. "62.313"

.Meas:

Measured value in mV at the indicator electrode associated to the cycle

number, e.g. "104.2".

.dWaterdt: Rate or drift associated to the cycle number in ug/min, e.g. "23.0".

.I:

Total charge in mA·s associated to the cycle number, e.g. "667.48".

.Pot:

Voltage at generator electrode associated to the cycle number.

0 means "undefined", 1 means <14 V, 2 means 14...28 V, 3 means

>28 V.

.IPulse:

Current of actual pulse associated to the cycle number.

1 means 100 mA, 2 means 200 mA, 3 means 400 mA.

OV will be sent for "overrange".

The unit "titrator" must be preset (see 4.2.2.92).

4.2.2.95. Setup.AutoInfo.Status ON, OFF

Setup.AutoInfo.P ON, OFF

Setup.AutoInfo.T.R ON, OFF

Setup.AutoInfo.T.G ON, OFF

Setup.AutoInfo.T.GC ON, OFF