Metrohm 756 KF Coulometer User Manual
Page 86

4.2 Remote control commands
756/831 KF Coulometer, Instructions for Use
78
4.2.2.27. Mode.CFmla
Mode.CFmla.1.Value
0...
±
999 999
Mode.CFmla.2.Value
etc., up to .19
Calculation constants specific to a method. Stored in the method memory of the
Coulometer. Operands specific to the sample (4.2.2.57 and 4.2.2.59) and values of
common variables (4.2.2.55) on the other hand are not stored with the methods.
4.2.2.28. UserMeth.FreeMem read
only
Memory space, available for user methods or silo lines. $Q sends the number of free
bytes, e.g.
"4928".
4.2.2.29. UserMeth.Recall $G
UserMeth.Recall.Name
up to 8 ASCII characters
UserMeth.Store $G
UserMeth.Store.Name
up to 8 ASCII characters
UserMeth.Delete $G
UserMeth.Delete.Name
up to 8 ASCII characters
UserMeth.DelAll $G
Management of the internal method memory: Load, store and delete methods. An
action is performed if "$G" is sent to the corresponding node just after entering the
name.
Do not use blank characters before and after method name!
.DelAll:
Deletes all methods in the user memory.
4.2.2.30. UserMeth.List.1.Name read
only
UserMeth.List.1.Mode read
only
UserMeth.List.1.Bytes read
only
UserMeth.List.1.Checksum read
only
for each method
List of the methods in the user method memory with the following characteristics:
.Name:
Name of the method
.Mode: Mode
.Bytes:
Number of bytes of the user memory used by the method
.Checksum: Checksum of the method, see 4.2.2.65.
4.2.2.31. Config.Monitoring.Reagent.Status ON,
OFF
Config.Monitoring.Reagent.Determ 1...99...999, OFF
Config.Monitoring.Reagent.DCounter
0...999
Config.Monitoring.Reagent.MaxTime 1...7...9999, OFF
Config.Monitoring.Reagent.TCounter
0...9999
Config.Monitoring.Reagent.ReagCap 1...1000...9999, OFF
Config.Monitoring.Reagent.RCounter
0...9999
Config.Monitoring.Reagent.ClearCount $G
Config.Monitoring.Reagent.Drift 0...99,
OFF