Metrohm 756 KF Coulometer User Manual
Page 85

4.2 Remote control commands
756/831 KF Coulometer, Instructions of Use
77
Mode.Def.Formulas.1.Decimal 0...2...5
Mode.Def.Formulas.1.Unit
6 ASCII characters
Mode.Def.Formulas.1.Limits ON,
OFF
Mode.Def.Formulas.1.LoLim
0...
±
999 999
Mode.Def.Formulas.1.UpLim
0...
±
999 999
Mode.Def.Formulas.1.Output active,
pulse,
OFF
Mode.Def.Formulas.2.Formula
etc. up to .9
Entry of formulas. Rules for formula entry, see page 34.
Example: "H2O*C01/C00"
In addition to the formula, a text for result output, the number of decimal places and
a unit for the result output can be selected. "No unit" is selected with the blank string.
In place of "RSX", a result name may be entered (.TextRS). This name is outputted in
the result report, scalc full and scalc srt. It is used for the result and the correspond-
ing mean value.
The limit control for results can also be activated. If a result is out of limit, a message
appears in the result report, E196 is sent, and output line L13 can be set.
4.2.2.23. Mode.Def.SiloCalc.Assign.C24
RSX, H2O, CXX
Mode.Def.SiloCalc.Assign.C25
RSX, H2O, CXX
Mode.Def.SiloCalc.MatchId
id1, id1&2, all, OFF
.Assign.C2X: Assignment to store results in the silo as C2X.
.MatchId:
Indication which sample identification(s) have to match so that the
results can be combined.
4.2.2.24. Mode.Def.ComVar.C30
RSX, MNX, H2O, CXX
Mode.Def.ComVar.C31
etc., up to .C39
Assignment of common variables.
The values of the common variables are to be found in &Config.ComVar. They can
be viewed and entered there, see 4.2.2.54.
4.2.2.25. Mode.Def.Report.Internal (only 756) result, water crv, rate crv,
meas crv, comb, mplist, param, calc, scalc full, scalc srt, ff
Mode.Def.Report.Assign1 ditto
Mode.Def.Report.Assign2
ditto
Definition of the report sequence, which is outputted automatically at the end of the
determination. Entries of more than one block have to be separated with ";".
.Internal: Internal printer of the Coulometer. (only 756)
.Assign1: Output to COM1 of the Coulometer.
.Assign2: Output to COM2 of the Coulometer.
4.2.2.26. Mode.Def.Mean.1.Assign RS1, RSX, H2O, CXX
Mode.Def.Mean.2.Assign
etc., up to .9
Assignment of the statistics calculations. Valid assignments are a requirement for
statistics calculations. In addition, the statistics calculation must be switched on, see
4.2.2.13. Rules for statistics calculations see page 37.