4.2 Remote control commands

756/831 KF Coulometer, Instructions for Use

80

IBM means the IBM character set following character set table 437 and IBM graphics

control characters. Select 'IBM' for work with the computer.

4.2.2.37. Config.PeriphUnit.RepToComport 1,2,1&2. And at 756: int.,1&int.,

2&int.,all. Selection of target for manually triggered reports.

int. Internal

printer.

1: COM1

2: COM2

4.2.2.38. Config.PeriphUnit.Balance Sartorius,Mettler,Mettler AT,

AND,Precisa

Selection of the balance type.

4.2.2.39. Config.PeriphUnit.Stirrer ON, OFF

Automatic stirrer control. With "ON" the stirrer will be switched on after starting of

conditioning. In the inactive state, the stirrer is switched off again.

4.2.2.40. Config.PeriphUnit.RemoteBox.Status

ON, OFF

Config.PeriphUnit.RemoteBox.Keyboard US, deutsch, francais,

español,

schweiz.

Config.PeriphUnit.RemoteBox.Barcode input, method, id1, id2,

id3, smpl size

Connections via Remote Box.

.Status:

Select if a Remote Box is connected.

.Keyboard: Type of keyboard which is connected to the Remote Box.

.Barcode: Select target in Coulometer where you wish to have the string from the

barcode reader. "input" means that the string comes into the field where

the cursor is currently placed.

4.2.2.41. Config.Aux.Language english, deutsch, francais, espanol,

italiano, portugese, svenska

Selection of the dialog language.

4.2.2.42. Config.Aux.Set $G

Config.Aux.Set.Date YYYY-MM-DD

Config.Aux.Set.Time HH:MM

Date and time.

Input format of the date: Year-month-day, two-digit, enter leading zeros.

Input format for the time: Hours:minutes, two-digit, enter leading zeros.

Date and time have to be set with &Config.Aux.Set $G just after entry of the value.

4.2.2.43. Config.Aux.RunNo

0...9999

Current sample number.

Set to 0 on power on and initialization. After 9999, counting starts again at 0.