4.1 General rules
756/831 KF Coulometer, Instructions for Use
58
RS send errors:
E42
CTS=OFF No proper handshake for more than 1 s.
Exit:
E43
The transmission of the KF Coulometer has been interrupted with XOFF for
at least 6 s.
Exit: Send XON or
E45
The receive buffer of the KF Coulometer contains an incomplete command
(L
F
missing). Sending from the KF Coulometer is therefore blocked.
Exit: Send L
F
or
E120
Overrange of the measured value.
Exit: Correct error or &m $S.
E121
Measuring point list overflow (more than 500 measuring points).
Exit: The error message disappears on next start.
E123
Missing EP for calculation.
Exit: The error message disappears on next start or on recalculation.
E127
Stop time reached.
Exit: The error message disappears on next start.
E128
No new mean.
Exit: The error message disappears on next start or on recalculation.
E129
No new common variable, old value remains.
Exit: The error message disappears on next start or on recalculation.
E132
Silo empty and it has been started with open silo or empty silo has been
opened.
Exit: Send a silo entry.
E133
Silo full.
Exit: Send new command.
E134
No method. A method, which is required from the silo memory, does not exist.
Exit: The error message disappears on next start.
E137
XXX Bytes are missing so that the method, the silo line could not be stored.
Exit: Send new command.
E155
No new silo result (C24 or C25).
Exit: The error message disappears on next start or on recalculation.
E176
The function &Assembly.Buret.Prep or &Assembly.Buret.Empty was interrupted
manually.
Exit: The error message disappears on next start.
E190
Overtitrated. The KF Coulometer is in the Iodine range.
Exit: The error message disappears when the Coulometer is again in the water range
or on next start.
E192
Check generator electrode: Not sufficient solvent in titration vessel or you are
working with fixed generator current or generator electrode defective. The results of a
determination may be erroneous and in the report you will find the message "
work.conditions not ok".
Exit: Rectify error.
E194
Sample unfit. Sample releases oxidative agents during titration.
Exit: Rectify error or &m $S.
E196
Result is out of limits.
Exit: The error message disappears on next start or on recalculation.