4.2 Remote control commands
756/831 KF Coulometer, Instructions of Use
81
4.2.2.44. Config.Aux.OpLevel standard, expert
Operator level for manual operation.
4.2.2.45. Config.Aux.StartDelay
0...999 999
Start delay time in s. During this time, the data of the preceding determination are re-
tained.
4.2.2.46. Config.Aux.ResDisplay bold, standard
Character set for the result display at the end of the determination.
4.2.2.47. Config.Aux.DevName
up to 8 ASCII characters
Name of the instrument for connections with several units. It is advisable to use only
the letters A...Z (ASCII No. 65...90), a...z (ASCII No. 97...122) and the numbers 0...9
(ASCII No. 48...57) when the function Setup.AutoInfo (4.2.2.97) is used at the same
time.
If a name has been entered, it will be printed out in the result report (full, short).
4.2.2.48. Config.Aux.Beep
1...3, OFF
Number of beep sounds.
4.2.2.49. Config.Aux.DisplayMeas ON,
OFF
Display of potentials during conditioning and titration.
4.2.2.50. Config.Aux.Prog read
only
Output of the program version.
The Coulometer sends "5.756.0010" on requests with $Q.
4.2.2.51. Config.RSSet1 $G
Config.RSSet1.Baud
300, 600, 1200, 2400, 4800, 9600
Config.RSSet1.DataBit 7,
8
Config.RSSet1.StopBit
1, 2
Config.RSSet1.Parity even,
odd, none
Config.RSSet1.Handsh HWs, SWchar, SWline, none
$G sets all RS settings. The changes are performed only if the instrument is inactive.
After the setting of the interface parameters, wait at least 2 s to allow the compo-
nents to equilibrate.
Settings of the values for the data transmission via the RS interface: baud rate, data
bit, stop bit, parity and type of handshake, see also page 97 ff.
The setting of the values must be initiated with $G immediately after entry of the val-
ues.
4.2.2.52. Config.Report.Id
ON, OFF
Config.Report.Instr
ON, OFF