4.2 Remote control commands

756/831 KF Coulometer, Instructions for Use

82

:

Report configuration. If a report line is switched off, the corresponding line will not

be outputted in the reports.

With "Run" on "OFF", only the run number will not be outputted, date (and time) are

available.

4.2.2.53. Config.ComVar.C30

with up to .C39, etc.

0...

±

999 999

Values of the common variables from C30 up to C39. Insert the common variables

directly or describe the determination results directly from the method, see 4.2.2.24.

4.2.2.54. SmplData.Status ON, OFF

On/off switching of silo memory. When the silo memory is switched on, the sample

data are fetched from the lowest valid silo line.

4.2.2.55. SmplData.OFFSilo.Id1

up to 12 ASCII characters

SmplData.OFFSilo.Id2

up to 12 ASCII characters

SmplData.OFFSilo.Id3

up to 12 ASCII characters

SmplData.OFFSilo.ValSmpl

6-digits, sign and decimal point

SmplData.OFFSilo.UnitSmpl

up to 5 ASCII characters

SmplData.OFFSilo.Limits read

only

Current sample data.

The identifications Id1...Id3 can be used in formulas as sample-specific calculation

constants C21...C23.

If "no unit" is desired for the unit of the sample size, the blank string must be entered.

.Limits:

Limits of sample size of current method.

4.2.2.56. SmplData.ONSilo.Counter.MaxLines read

only

SmplData.ONSilo.Counter.FirstLine read

only

SmplData.ONSilo.Counter.LastLine read

only

Information on silo memory.

.MaxLines: Maximum possible number of silo lines.

.FirstLine: Lowest valid silo line.

.LastLine: Last occupied silo line.

4.2.2.57. SmplData.ONSilo.EditLine.1.Method

up to 8 ASCII characters

SmplData.ONSilo.EditLine.1.Id1

up to 12 ASCII characters

SmplData.ONSilo.EditLine.1.Id2

up to 12 ASCII characters

SmplData.ONSilo.EditLine.1.Id3

up to 12 ASCII characters

SmplData.ONSilo.EditLine.1.ValSmpl 6-digits, sign and dec.point

SmplData.ONSilo.EditLine.1.UnitSmpl

up to 5 ASCII characters

SmplData.ONSilo.EditLine.1.C24 read

only

SmplData.ONSilo.EditLine.1.C25 read

only

SmplData.ONSilo.EditLine.1.Mark read

only

etc., up to .255

Contents of a silo line.