Config – Metrohm 756 KF Coulometer User Manual
Page 72
4.2 Remote control commands
756/831 KF Coulometer, Instructions for Use
64
&Config
Object Description
Input
range
Reference
& Root
:
├
Config
Instrument configuration
:
├
.Monitoring
Monitoring functions
│ ├
.Reagent
Monitoring of reagent
4.2.2.31.
│ │ ├
.Status
Status of reagent monitoring
ON, OFF
ditto
│ │ ├
.Determ
Number of determinations
1...99...999, OFF
ditto
│ │ ├
.DCounter Determination
counter
0...999 ditto
│ │ ├
.MaxTime Time
monitoring
1...7...9999, OFF
ditto
│ │ ├
.TCounter Time
counter
0...9999 ditto
│ │ ├
.ReagCap
Reagent capacity monitoring
1...1000...9999, OFF
ditto
│ │ ├
.RCounter
Reagent capacity counter
0...9999 ditto
│ │ ├
.ClearCount
Clears all counters above
$G
ditto
│ │ ├
.Drift
Change if drift is above
0...99, OFF ditto
│ ├
.Change
Change of reagent
$G, $S
4.2.2.32.
│ │ ├
.Status
Type of reagent changing
auto, man., OFF
ditto
│ │ ├
.WaitTime
Waiting after switching off stirrer
0...999 999
ditto
│ │ ├
.AspVol
Volume to aspirate
0...100...9999 ditto
│ │ ├
.SolventVol
Solvent volume to add
0...100...9999 ditto
│ │ ├
.Rinse Rinsing
volume
0...9999 ditto
│ │ ├
.NoRinse
Number of rinsing cycles
1...9 ditto
│ ├
.Validation Validation
monitoring
4.2.2.33.
│ │ ├
.Status
Status of validation monitoring
ON, OFF
ditto
│ │ ├
.Interval
Time interval for validation
1...365...9999 ditto
│ │ ├
.Counter Time
counter
0...9999 ditto
│ │ ├
.ClearCount
Clears the counter above
$G
ditto
│ ├
.Service
Monitoring of Metrohm service
$G
4.2.2.34.
│ │ ├
.Status
Status of service monitoring
ON, OFF
ditto
│ │ ├
.Date
Date of next service
XXXX-XX-XX
ditto
│ ├
.DiagRep
Printing of system test report
ON, OFF 4.2.2.35.
├
.PeriphUnit
Selection of peripheral units
│ ├
.CharSet1
External printer on COM1
Epson,Seiko,Citizen
│ │
IBM,HP 4.2.2.36.
│ ├
.CharSet2
as for CharSet1
│ ├
.RepToComport Output of manual reports
1, 2, 1&2. And in 756:
int.,1&int.,2&int.,all
4.2.2.37.
│ ├
.Balance
Selection of balance
Sartorius,Mettler,Mettler AT
│ │
AND,Precisa 4.2.2.38.
│ ├
.Stirrer Stirrer
control
ON, OFF
4.2.2.39.
│ ├
.RemoteBox
Connected remote box
4.2.2.40.
│ │ ├
.Status Status
ON,
OFF ditto
│ │ ├
.Keyboard
Type of keyboard
US, deutsch, francais,
│ │ │
español,
schweiz.
ditto
│ │ ├
.Barcode
Input of barcode reader
input, method, id1, id2,
│ │ │
id3, smpl size
ditto
├
.Aux
Miscellaneous
│ ├
.Language Dialog
language
english, deutsch,
│ │
francais, espanol, italiano,
│ │
portugese, svenska
4.2.2.41.