beautypg.com

5 appendix, Warranty, Eu declaration of conformity – Metrohm 754 Dialysis Unit User Manual

Page 38: Certificate of conformity and system validation, 5appendix, 1 technical data

background image

5.1 Technical data

754 Dialysis Unit

35

5

Appendix

5.1

Technical data

Dialysis cell

Material

Plexiglass (polymethyl methacrylate)

Solvent compatibility

Water, ethanol

(no organic sovents!)

Cell volume

240

µ

L (from inlet to outlet opening)

Diameter of membrane

47 mm

Pump

Pump type

2-channel peristaltic pump with rotational speed

of

20/min (50 Hz)

24/min (60 Hz)

Pump capacity

(with water,

without counterpressure)

with 6.1826.030 pump tubing:

typically 0.5

0.6 mL/min

with 6.1826.040 pump tubing:

typically 0.9

1.1 mL/min

Pressure

max. 1.5 bar (0.15 MPa)

Pump tubing material

PVC (Tygon

)

Mains connection

Mains voltage

115 V: 100...120 V

±

10 %

230 V: 220...240 V

±

10 %

Switchable with voltage selector in fuse holder

(see section 2.2.1)

Mains frequency

50...60 Hz

Power consumption

12 VA

Fuse

5 mm Ø, 20 mm length

100

120 V: 0.2 A (slow-blow)

220

240 V: 0.1 A (slow-blow)

Remote interface

Purpose

Remote control of pump (switching on/off)

Pin assignment

330

+5V

2

5

+

-

Pin

Current flows:

on

Current does not flow:

off

external

754