Metrohm 754 Dialysis Unit User Manual
Page 30

3.3 The dialysis procedure
754 Dialysis Unit
27
2 Measure the standard with dialysis
•
Set the dialysis time in the program to 5 min (see section
3.3.2).
•
Immerse sample suction tubing 26
26 in the standard used in
point 1.
•
Start the program on the 732 IC Detector with the R/S> key. Wait until the LED in the flashing, signalling the end of the evaluation. 3 Additional measurements • Increase the dialysis time in the program (see section 3.3.2) by 5 min increments until the measured values remain con- stant. • Take measurements as per point 2. 4 Determine optimum dialysis time • Record the ratio of the peak surfaces, i.e. concentrations measured with and without dialysis in relation to the dialysis time (i.e. the recovery rate) and determine the optimum dialy- sis time. 3.3.6 Setting the program time Set the program time so that the chromatogram can be completely re- corded and evaluated within this period. 3.3.7 Dialysis procedure To analyse samples using dialysis you should preferably proceed as follows: 1 Prepare acceptor solution • To avoid malfunctions caused by air bubbles in the acceptor channel, the ultrapure water used for the acceptor solution should always be degassed for at least 10 min with vacuum, N 2 or He. 2 Prepare sample • To prevent blockages in the sample channel, samples con- taining a high volume of (suspended/solid particles) should always be centrifuged using a table centrifuge at 10,000 rev./min for 5 min.