3.3 The dialysis procedure

754 Dialysis Unit

27

2 Measure the standard with dialysis

Set the dialysis time in the program to 5 min (see section

3.3.2).

Immerse sample suction tubing 26

26 in the standard used in

point 1.

Start the program on the 732 IC Detector with the

R/S> key. Wait until the LED in the key stops

flashing, signalling the end of the evaluation.

3 Additional measurements

Increase the dialysis time in the program (see section 3.3.2)

by 5 min increments until the measured values remain con-

stant.

Take measurements as per point 2.

4 Determine optimum dialysis time

Record the ratio of the peak surfaces, i.e. concentrations

measured with and without dialysis in relation to the dialysis

time (i.e. the recovery rate) and determine the optimum dialy-

sis time.

3.3.6

Setting the program time

Set the program time so that the chromatogram can be completely re-

corded and evaluated within this period.

3.3.7

Dialysis procedure

To analyse samples using dialysis you should preferably proceed as

follows:

1 Prepare acceptor solution

To avoid malfunctions caused by air bubbles in the acceptor

channel, the ultrapure water used for the acceptor solution

should always be degassed for at least 10 min with vacuum,

N

2

or He.

2 Prepare sample

To prevent blockages in the sample channel, samples con-

taining a high volume of (suspended/solid particles) should

always be centrifuged using a table centrifuge at 10,000

rev./min for 5 min.