2.3 Attaching the accessories
754 Dialysis Unit
17
2.3.5
Connecting the dialysis cell
To connect the dialysis cell to the IC system proceed as follows (see
Fig. 8):
1 Tubing connections for acceptor solution
•
Mount a PVDF compression fitting 50
50 (6.2744.000) on the
end of PEEK capillary 24
24 projecting from feedthrough 27
27 (see
Fig. 5).
•
Screw PEEK capillary 24
24 to inlet opening 42
42 of the bottom
part 35
35 of the cell (see Fig. 7).
•
Mount a PVDF compression fitting 50
50 (6.2744.000) to the end
of PEEK capillary 25
25 projecting from feedthrough 27
27 (see Fig.
5).
•
Screw PEEK capillary 25
25 to outlet opening 41
41 of the bottom
part 35
35 of the cell (see Fig. 7).
2 Tubing connections for sample
•
Mount a PVDF compression fitting 50
50 (6.2744.000) on the
end of the PTFE capillary 33
33 connected to pump tubing 30
30.
•
Screw PTFE capillary 33
33 to inlet opening 43
43 of the top part 38
38
of the cell (see Fig. 7).
•
Mount a PVDF compression fitting 50
50 (6.2744.000) to one
end of the PTFE capillary 45
45 (6.1803.040; L = 1 m).
•
Screw PTFE capillary 45
45 to outlet opening 44
44 of the top part
38
38 of the cell (see Fig. 7).
Do not connect the dialysis cell with anything other than the
6.2744.000 PVDF compression fittings supplied. If 6.2744.010
PEEK compression fittings are used, stress cracks may appear in the
dialysis cell !