beautypg.com

5 setting the dialysis time – Metrohm 754 Dialysis Unit User Manual

Page 29

background image

3 Operation

754 Dialysis Unit

26

0.0

2.0

4.0

6.0

8.0

10.0

12.0

0.0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1.0

Transfer time [min]

Concentration [mg/L]

Fluoride
Chloride
Nitrite
Bromide
Nitrate
Phosphate
Sulphate

3.3.5

Setting the dialysis time

After calculating the optimum transfer time, determine the optimum di-

alysis time. This value will depend on the total ion concentration. The

dialysis time with a stopped acceptor flow should be selected so that

100% of the sample concentration is achieved in the acceptor solution.

For the 6.2714.010 dialysis membrane (cellulose acetate; thick-

ness = 115

µ

m; pore size = 0.2

µ

m) available from Metrohm, the opti-

mum dialysis time for a total ion concentration of

5 mg/L is 10 min.

If a different dialysis membrane is used, or measurements are taken

with a total ion concentration of < 5 mg/L, then the optimum dialysis

time will have to be redetermined by measuring the recovery rate in re-

lation to time. Proceed as follows:

1 Measure the standard directly

Switch off remote control of the 754 Dialysis Unit with the

key 22 (see section 3.1).

Set the two valves A and B on the 733 IC Separation Center

to "FILL" with keys 23

23 and 25

25.

Immerse suction tubing 27

27 in the required ion standard,

which has approximately the same total ion concentration as

the samples.

Switch on the pump on the 754 Dialysis Unit with the

key 11.

After approx. 2 min, switch valve A on the 733 IC Separation

Center to "INJECT" with key 24

24 and simultaneously start the

data evaluation system. In operation with the 753 Suppressor

Module, the suppressor has to be rotated with the

key.

Immerse suction tubing 27

27 in the acceptor solution and rinse

the acceptor channel for approx. 2 min.

Switch off the pump on the 754 Dialysis Unit with the

key 11.

Switch the remote control of the 754 Dialysis Unit back on

with the key 22.