2.3 Attaching the accessories

754 Dialysis Unit

11

2 Connect new feedthrough

Using a wrench or a spanner, unscrew the nut on the interior

side of feedthrough 27

27 and remove the feedthrough.

Insert new 6.2756.000 tubing feedthrough in the hole and

retighten the nut.

3 Attach tubing to valve A

Screw 6.1825.210 PEEK sample loop (20

µ

L) to connections

"3" and "6" of valve A 18

18 using the enclosed 6.2744.010 PEEK

compression fittings.

Screw PEEK capillary 19

19 (6.1831.060; L =1 m) to connection

"2" of valve A 18

18 using a 6.2744.010 PEEK compression fit-

ting. Pull the other end of the tubing to the back through the

rear panel opening 41

41.

Screw PEEK capillary 20

20 (6.1831.040; L =15 cm) to connec-

tion "1" of valve A 18

18 using a 6.2744.010 PEEK compression

fitting.

4 Attach tubing to valve B

Screw the PEEK capillary 20

20 (6.1831.040; L =15 cm) at-

tached to connection "1" of valve A 18

18 to connection "3" of

valve B 22

22 using a 6.2744.010 PEEK compression fitting.

Screw PEEK capillary 21

21 (6.1831.040; L =15 cm) to connec-

tions "1" and "2" of valve B 22

22 using two 6.2744.010 PEEK

compression fittings.

Screw PEEK capillary 25

25 (6.1831.050; L =40 cm) to connec-

tion "4" of valve B 22

22 using a 6.2744.010 PEEK compression

fitting and pull the other end to the front through feedthrough

27

27.

Screw PEEK capillary 24

24 (6.1831.050; L =40 cm) to connec-

tion "5" of valve B 22

22 using a 6.2744.010 PEEK compression

fitting and pull the other end to the front through feedthrough

27

27.

Screw PEEK capillary 23

23 (6.1831.050; L =40 cm) to connec-

tion "6" of valve B 22

22 using a 6.2744.010 PEEK compression

fitting and pull the other end to the front through feedthrough

28

28.

5 Connecting to the 753 Suppressor Module (Option)

Insert the suppressor block in the 733 IC Separation Center

as per the 753 Instructions for Use.

Connect tubing to the IC column, the detector block and the

waste in the 733 IC Separation Center as per the 753 Instruc-

tions for Use.

Pull the two inlet capillaries for H

2

SO

4

and H

2

O to the front

through feedthrough 27

27 and connect to the 753 Suppressor

Module pump tubings.