3.3 The dialysis procedure

754 Dialysis Unit

23

3.3

The dialysis procedure

3.3.1

Dialysis flow chart

Every sample dialysed with the 754 Dialysis Unit comprises the follow-

ing four steps:

1 Rinsing

Valve A: FILL

Valve B: INJECT
Sample channel, ac-

ceptor channel and

sample loop are rinsed

(rinsing time: approx. 2

min).

A

B

Fill

Inject

2 Dialysis with «Stopped Flow»

Valve A: FILL

Valve B: FILL
The sample is dialysed

with resting acceptor

solution (dialysis time:

approx. 10 min).

A

B

Fill

Fill

3 Transfer to sample loop

Valve A: FILL

Valve B: INJECT
The acceptor solution

enriched with ions from

the sample is trans-

ferred to the sample

loop (transfer time:

approx. 0.4 to 0.6 min).

A

B

Fill

Inject

4 Injection

Valve A: INJECT

Valve B: INJECT
The acceptor solution

enriched with ions from

the sample is injected

into the IC system.

A

B

Inject

Inject

Eluent

Column

Waste

Sample

Acceptor

Eluent

Column

Waste

Sample

Acceptor

Eluent

Column

Waste

Sample

Acceptor

Eluent

Column

Waste

Sample

Acceptor