beautypg.com

Metrohm 754 Dialysis Unit User Manual

Page 28

background image

3.3 The dialysis procedure

754 Dialysis Unit

25

3.3.3

Setting the rinsing time

A rinsing time of 2 min is normally sufficient to ensure thorough rinsing

of the sample and acceptor channels. This time can be increased if re-

quired.

3.3.4

Setting the transfer time

The time allowed for transferring the enriched acceptor solution to the

sample loop must be set to ensure that the part of the acceptor solution

with the highest concentration of ions is transferred to the sample loop.

With standard accessories the optimum transfer time depends on the

rate of operation of the pump; the following values can be used as a

guideline:

Mains frequency Optimum transfer time

50 Hz

0.5 min

60 Hz

0.4 min

For each analysis problem, determine the optimum transfer time by

taking measurements of the individual ion concentrations in relation to

the transfer time itself and verify your measurements from time to time.

Proceed as follows:

1 First measurement

Set the transfer time in the program to 0.2 min and the dialy-

sis time to 10 min (see section 3.3.2).

Immerse sample suction tubing 26

26 in the standard with 10

mg/L of the required anion or cation.

Start the program on the 732 IC Detector with the

R/S> key. Wait until the LED in the key stops

flashing, signalling the end of the measuring process.

2 Additional measurements

Increase the transfer time in the program (see section 3.3.2)

by increments of 0.1 min until the concentration measured

falls again.

Take measurements as in point 1.

3 Determining the optimum transfer time

Record the measured peak area resp. concentration in rela-

tion to the transfer time and determine the optimum transfer

time.

The Figure below shows an example of this determination process for

anions using the 6.2714.010 dialysis membrane (cellulose acetate;

thickness = 115

µ

m; pore size = 0.2

µ

m) available from Metrohm. The

sample was a standard solution containing 10 mg/L of each of the indi-

vidual anions, which was dialysed for 10 min (mains frequency 50 Hz).