beautypg.com

3 attaching the accessories – Metrohm 754 Dialysis Unit User Manual

Page 13

background image

2 Installation

754 Dialysis Unit

10

2.3

Attaching the accessories

2.3.1

Electrical connection to the 732 IC Detector

In order to allow remote operation of the pump via the 732 IC Detector,

connection 15

15 of the 754 Dialysis Unit must be connected to the

"Remote" connection of the 732 IC Detector by means of the

6.2143.200 cable as per Fig. 4. The switch 22 on the front

of the 754 Dialysis Unit must additionally be set to "ON" to permit re-

mote operation.
If you wish to connect a second, additional unit (e.g. 752, 753, 754) to

the remote interface of the 732 IC Detector, connect the Dialysis Unit

754 using the optionally available 6.2143.220 cable instead of the

6.2143.200 cable.

6.2143.200

754

732

Fig. 4: Connection of 754 Dialysis Unit to 732 IC Detector

2.3.2

Attaching the accessories to the 733.0020 IC Separation
Center

The 2.733.0020 IC Separation Center comes ready to operate with two

IC columns. To operate it with the 754 Dialysis Unit, reassemble it as

follows (see Fig. 5 and Fig. 8):

1 Dismantle accessories

Unscrew the two sample loops mounted on valve A 18

18 and

valve B 22

22.

Loosen the rotary nipples screwed onto the interior side of

feedthrough 22

22 and 28

28, pull the two PTFE suction tubings

right out of the connections and unscrew from connection "1"

of valve A 18

18 and valve B 22

22.

Remove the PEEK capillaries running from connection "2" of

valve A 18

18 and valve B 22

22 to feedthroughs 21

21 and 27

27.