beautypg.com

Weidmuller WI-I/O 9-L: Wireless I/O Transmitter / Receiver v1.3 User Manual

Page 81

background image

WI-I/O 9-L Configuration Manual

WI-I/O 9-L Wireless I/O

page 81 of 108

&

&

This section describes how to set additional unit detail information including:

Sample period and warmup time

- the WI-I/O 9-L-T only turns on the analog circuits

and the +24V analog loop power supply when they are actually required. You can

reduce the module power consumption by limiting the amount of time the analog

circuitry and the +24V analog loop supply is enabled.

Sample period determines how often the module checks an analog signal. Warmup

time determines how long the analog circuits are powered up before performing the

sample. The analogue circuitry is powered all the time if the warmup period for any

analog input is set to a value longer than the sampling period.

By setting the sample time as long as possible, and setting the warmup time as short

as necessary to power the external conditioning circuitry, the module power

consumption can be reduced; and

Debounce time

- the amount of time an input needs to have differed from its previous

value before it is recognised as having changed. This helps with noisy signals.

Inputs subject to a high level of noise can cause the module to transmit multiple

messages due to random changes to the input level.

By setting the debounce time to a larger value, the signal needs to change by the

configured sensitivity for some time before a change is registered and a new message

is sent.

WI Series lets you set the following options:

Feature

Description

Example

Thermocouple debounce

For analog signals (e.g. thermocouple, analog

4-20 mA, and module temperature), debounce

time is the product of the number of samples

and the sample period. The signal must differ

from the previous value by the sensitivity value

for this duration to register a ‘change of state’.
This means that changing the analog sample

time will also change the debounce time. For

analog signals a value of 0 or 1 generates the

quickest response, and 255 the slowest

response.

Thermocouple sample

period

This sets the period between samples of the

thermocouple input.
A smaller value will cause mappings using this

input to be triggered more frequently (i.e.

quicker response).
A larger value will cause mappings using this

input to be triggered less frequently (i.e. less

power consumption).