Muster – Memmert PM 200/400/500 Pump Module (Generation 2003) User Manual

Page 41

background image

VO

Muster

h

mB

mB

h

mb

mB

h

mB

mB

page 41

Note:

When all the calibration corrections are set to 0mb, the oven is restored to the factory

calibration

Calibration correction

Range –10mB to +10mb

Vacuum calibration

adjustment range: down to

20 mbar below CAL2

Calibration point 1

Calibration correction

Range –10mB to +10mb

Vacuum calibration

adjustment range: 20 mbar

above CAL1 to 20 mbar

below CAL3

Calibration point 2

Calibration correction

Range –10mB to +10mb

Vacuum calibration adjustment

range: 20 mbar above CAL2 up

to 1100 mbar

Calibration point 3

25 Communication interface for the PC

25.1 USB interface

The chamber is fitted by default with a USB interface in accordance with the USB specification. With this
interface it is possible to control and log the chamber remotely from the computer. For this, the “Celsius“
software is used.
The chamber must be given a unique device address in the SETUP submenu, menu item ADDRESS, via which
the computer communicates with the oven . The default setting is ADDRess 0. Using this, the appropriate
oven can be selected and programmed from the computer.
If several chambers are to be connected to a computer via USB interface, an appropriate interface on the
computer and a separate cable are required for each chamber.
The maximum cable length is 5 m.