Note:
When all the calibration corrections are set to 0mb, the oven is restored to the factory
calibration
Calibration correction
Range –10mB to +10mb
Vacuum calibration
adjustment range: down to
20 mbar below CAL2
Calibration point 1
Calibration correction
Range –10mB to +10mb
Vacuum calibration
adjustment range: 20 mbar
above CAL1 to 20 mbar
below CAL3
Calibration point 2
Calibration correction
Range –10mB to +10mb
Vacuum calibration adjustment
range: 20 mbar above CAL2 up
to 1100 mbar
Calibration point 3
25 Communication interface for the PC
25.1 USB interface
The chamber is fitted by default with a USB interface in accordance with the USB specification. With this
interface it is possible to control and log the chamber remotely from the computer. For this, the “Celsius“
software is used.
The chamber must be given a unique device address in the SETUP submenu, menu item ADDRESS, via which
the computer communicates with the oven . The default setting is ADDRess 0. Using this, the appropriate
oven can be selected and programmed from the computer.
If several chambers are to be connected to a computer via USB interface, an appropriate interface on the
computer and a separate cable are required for each chamber.
The maximum cable length is 5 m.