23.2.4 Automatic temperature monitor (ASF)

AUTO

A monitoring device which automatically follows the selected temperature setpoint.
The tolerance margin of the ASF is set in SETUP

(see the menu item Tolerance margin ASF in the Section

„Basic oven settings SETUP“).

Automatic temperature
monitor OFF

(ASF OFF)

MIN

MAX

AUTO

°C

Using the push/turn control select the AUTO
symbol.
Hold down the SET key and select off using
the push/turn control.

Automatic temperature
monitor ON

(ASF ON)

MIN

MAX

AUTO

°C

Using the push/turn control select the AUTO
symbol.
Hold down the SET key and select on using
the push/turn control.

Notes on the ASF:

The tolerance margin for the ASF is selected in SETUP

(see the menu item Tolerance margin ASF in the

Section „Basic oven settings SETUP”).

Tolerance margin reached = ASF activated

The ASF is automatically activated when the actual temperature has reached 50% of the selected tolerance
margin of the setpoint (in the example 150°C – 5°C).
The activation of the automatic temperature monitor is indicated by the bright

AUTO

-symbol.

160°C

140°C

ASF alarm

ASF activated

ASF activated

A

ASF set to

+/-10°C in

SETUP