8. Setting the air/gas connection IN1 of the first ramp
segment (only with corresponding additional fittings)
Turn the push/turn control to the right until the INLET display flashes.
Hold down the SET key and select the desired air/gas connection IN1
with the push/turn control.
9. Set closure command of first ramp segment
Turn the push/turn control clockwise until a segment closure command
(e.g. end ) appears.
Hold down the SET key and set SPWT [T] with the push/turn control.
10. Select duration of second ramp segment
Using the push/turn control select the time indication.
Hold down the SET key and set the time 00:01 using the push/turn
control.
h
11. Select temperature of second ramp segment
Turn the push/turn control clockwise until the temperature display is
flashing.
Hold down the SET key and set the required temperature setpoint of
180°C
using the push/turn control.
°C
12. Setting the vacuum of second ramp segment
Rotate push/turn control clockwise until the vacuum indication is
flashing.
Hold down the SET key and use the push/turn control to set the required
vacuum of 50mb.
mB
13. Setting the air/gas connection IN1 of the second ramp
segment (only with corresponding additional fittings)
Turn the push/turn control to the right until the INLET display flashes.
Hold down the SET key and select the desired air/gas connection IN1
with the push/turn control.
14. Set closure command for second ramp segment
Turn the push/turn control clockwise until a segment closure command
(e.g end ) appears.
Hold down the SET key and set SPWT (P) with the push/turn control.
15. Select duration of third ramp segment
Using the push/turn control select the time indication
Hold down the SET key and set the time 02:00 using the push/turn
control.
h
16. Select temperature of third ramp segment
Turn the push/turn control clockwise until the temperature display is
flashing.
Hold down the SET key and set the required temperature setpoint of
180°C
using the push/turn control.
°C
17. Setting the vacuum of third ramp segment
Rotate push/turn control clockwise until the vacuum indication is
flashing.
Hold down the SET key and use the push/turn control to set the required
vacuum of 50mb.
mB