The manually set monitor temperature and the electronic overtemperature protection are monitored on
VO-ovens by an adjustable temperature monitor (TWW) Protection Class 3.1 to DIN 12 880:

23.2.3 Adjustable temperature monitor (TWW) Protection Class 3.1 to DIN 12 880
If the manually set monitor temperatur MAX is exceeded, the TWW takes over the control of the

temperature and starts to control at the monitor temperature. As a warning the alarm symbol

is

flashing.

TB approx. 20°C above Tmax

TWW set manually

setpoint temperature

t

°C

controller fault

emergency operation