20.1 Closure commands for ramp segments

NEXT

Follow-on with next programme segment.

SPWT (T)

SET-POINT WAIT

Wait until the setpoint temperature is reached.
The oven only starts the next programme segment when the programmed
setpoint temperature has been reached, even if the programmed
heating time has already elapsed.

SPWT (P)

SET-POINT WAIT

Wait until the set vacuum has been reached.
The oven starts the next programme segment only when the programmed
vacuum has been reached, even if the programmed segment time has
already elapsed or if the set temperature has already been reached.

SPWT (PT)

SET-POINT WAIT

Wait until temperature setpoint and vacuum setpoint have been
reached.
The oven starts the next programme segment only when the programmed
setpoint temperature and the programmed vacuum have been reached,
even if the programmed segment time has already elapsed.

LOOP

Ramp repeat function.
The set programme is repeated after passing through all programmed
segments.
1-99

= repeats

cont

= continuous repeat function

HOLD

End of programme without switching off the heating; temperature and
all other settings (e.g. air valve) are maintained.

END

Programme end, heating is switched off, chamber is vented.

The programme segments are linked together by the segment closure command. These commands
therefore control the programme sequence.

°C

Ramp

closure

command

ramp

segment 1

spwt (t)

segment 1

closure

command

ramp

segment 2

next

closure

command

ramp

segment 4

next

closure

command

ramp

segment 5

end

closure

command

ramp

segment 3

spwt (t)

segment 3

segment 1

segment 5

segment 4

delayed

programme start