Muster, Ramp – Memmert PM 200/400/500 Pump Module (Generation 2003) User Manual
Page 25
VO
Muster
page 25
20.1 Closure commands for ramp segments
NEXT
Follow-on with next programme segment.
SPWT (T)
SET-POINT WAIT
Wait until the setpoint temperature is reached.
The oven only starts the next programme segment when the programmed
setpoint temperature has been reached, even if the programmed
heating time has already elapsed.
SPWT (P)
SET-POINT WAIT
Wait until the set vacuum has been reached.
The oven starts the next programme segment only when the programmed
vacuum has been reached, even if the programmed segment time has
already elapsed or if the set temperature has already been reached.
SPWT (PT)
SET-POINT WAIT
Wait until temperature setpoint and vacuum setpoint have been
reached.
The oven starts the next programme segment only when the programmed
setpoint temperature and the programmed vacuum have been reached,
even if the programmed segment time has already elapsed.
LOOP
Ramp repeat function.
The set programme is repeated after passing through all programmed
segments.
1-99
= repeats
cont
= continuous repeat function
HOLD
End of programme without switching off the heating; temperature and
all other settings (e.g. air valve) are maintained.
END
Programme end, heating is switched off, chamber is vented.
The programme segments are linked together by the segment closure command. These commands
therefore control the programme sequence.
°C
Ramp
closure
command
ramp
segment 1
spwt (t)
segment 1
closure
command
ramp
segment 2
next
closure
command
ramp
segment 4
next
closure
command
ramp
segment 5
end
closure
command
ramp
segment 3
spwt (t)
segment 3
segment 1
segment 5
segment 4
delayed
programme start