Memmert PM 200/400/500 Pump Module (Generation 2003) User Manual
Page 28
![background image](/manuals/567639/28/background.png)
VO
page 28
18. Setting the air/gas connection IN1 of the third ramp
segment (only with corresponding additional fittings)
Turn the push/turn control to the right until the INLET display flashes.
Hold down the SET key and select the desired air/gas connection IN1
with the push/turn control.
19. Set closure command for third ramp segment
Turn the push/turn control clockwise until a segment closure command
(e.g end ) appears.
Hold down the SET key and set NEXT with the push/turn control.
20. Select duration of fourth ramp segment
Using the push/turn control select the time indication
Hold down the SET key and set the time 02:00 using the push/turn
control.
h
21. Select temperature of fourth ramp segment
Turn the push/turn control clockwise until the temperature display is
flashing.
Hold down the SET key and set the required temperature setpoint of
30°C
using the push/turn control.
°C
22. Setting the vacuum of fourth ramp segment
Rotate push/turn control clockwise until the vacuum indication is
flashing.
Hold down the SET key and use the push/turn control to set the required
vacuum of 1050mb.
mB
23. Setting the air/gas connection IN1 of the fourth ramp
segment (only with corresponding additional fittings)
Turn the push/turn control to the right until the INLET display flashes.
Hold down the SET key and select the desired air/gas connection IN1
with the push/turn control.
24. Set closure command for fourth ramp segment
Turn the push/turn control clockwise until a segment closure command
(e.g end ) appears.
Hold down the SET key and set END with the push/turn control.
25. Exit programme writing mode EDIT
Turn the push/turn control clockwise until EXIT appears on the display.
Press the SET key briefly to enter.
26. Set temperature monitor
Turn the push/turn control clockwise and set the temperature monitor.
(see Section: “Temperature monitor”)
MIN
MAX
AUTO
°C
27. Start programme
Turn the push/turn control anticlockwise until the stop symbol
is
flashing.
Hold down the SET key and select Start
with the push/turn control.