Vo page 24, H°c rh mb – Memmert PM 200/400/500 Pump Module (Generation 2003) User Manual
Page 24

VO
page 24
SETUP
loop
t3
t4
t2
t1
on
off
Mo
Tu
We
Th
Fr
Sa
Su
MAX
°C
h
°C
rh
mb
%
IN 1
IN 2
OUT
Duration of first ramp segment
Range: 1 minute to 999 hours
SETUP
loop
t3
t4
t2
t1
on
off
Mo
Tu
We
Th
Fr
Sa
Su
MAX
°C
h
°C
rh
mb
%
IN 1
IN 2
OUT
Setpoint temperature / temperature at the end of the ramp segment
Range: 20°C to 200°C
SETUP
loop
t3
t4
t2
t1
on
off
Mo
Tu
We
Th
Fr
Sa
Su
MAX
°C
h
°C
rh
mb
%
IN 1
IN 2
OUT
Vacuum at the end of the ramp segment
Adjustment range: 10 to 1100 mbar or LO
SETUP
loop
t3
t4
t2
t1
on
off
Mo
Tu
We
Th
Fr
Sa
Su
MAX
°C
h
°C
rh
mb
%
IN 1
IN 2
OUT
Closure command of ramp segment
Range: next, spwt, loop, hold, end
(see Section: „Closure commands for ramp segments“)
SETUP
loop
t3
t4
t2
t1
on
off
Mo
Tu
We
Th
Fr
Sa
Su
MAX
°C
h
°C
rh
mb
%
IN 1
IN 2
OUT
Exit the programme write mode EDIT
Turn the push/turn control clockwise unti EXIT appears on the display, briefly press the SET key to enter