Setpoint changed = ASF de-activated automatically

If the temperature setpoint is altered, the automatic temperature monitor is automatically de-activated
temporarily (see in the example the setpoint is changed from 150°C to 75°C) until the tolerance margin
of the new temperature setpoint is reached (see in the example below: the ASF is re-activated at 75°C +/-
10°C).

160°C

140°C

ASF alarm

ASF activated

ASF activated

160°C

140°C

ASF-alarm

ASF activated

ASF activated

85°C

65°C

ASF set to

+/-10°C in

SETUP