21 Printer

PRINT

PRINT

SETUP

VO ovens may be equipped with a parallel printer interface, as used in personal computers.

This parallel printer interface on the back of the oven is suitable for connecting conventional PCL3-
compatible ink jet and laser printers which are provided with a parallel printer interface (e.g. HP Deskjet
5550 or HP Deskjet 9xx).

It is important to use a screened interface cable. The screen must be connected to the plug case.

The controller is provided with an internal log memory

(see Section: „Log memory”).

The report data can

in this operating mode be printed out through the printer connected to the oven.

When using a colour printer, the various graphics can be printed in colour.

On the printout the GLP data head is also printed automatically and contains the following information:
- Printing

date

-

Time period of report

-

Running page number

-

Serial number and oven designation

By turning the push/turn control the following parameters can be selected in turn and altered as described
in the

Section Setting the parameters.

Reading the date of the first print page

Reading the date of the last print page

Start graphics print

Print programme and configuration page

Exit print menu and back to main menu