Select operating mode “normal operation, Select temperature setpoint, Select monitor temperature – Memmert PM 200/400/500 Pump Module (Generation 2003) User Manual
Page 20: Setting the vacuum

VO
page 20
Setting example “Normal operation“
The oven (VO500) has to heat up to 180°C and be evacuated to 50 mbar.The monitoring function has to
operate at 200°C.
Temperature
in °C
Pressure
in mb
t
t
180°C
1. Select operating mode “Normal operation“
After holding down the SET key (approx. 3 sec), the current operating
mode is flashing. Select operating mode I with the push/turn control
while holding down the SET key.
After the SET key has been released the controller is in
operating mode I.
SETUP
2. Select temperature setpoint
Hold down the SET key and use the push/turn control to select the
required temperature setpoint of 180°C.
After the SET key has been released the oven briefly flashes the
temperature setpoint.
The display then changes to the actual temperatureand the controller
starts to control to the selected temperature setpoint 180°C.
Heating is indicated by the orange heater symbol.
°C
5. Select monitor temperature
Turn the push/turn control clockwise until the overtemperature display
MAX
is flashing. Hold down the SET key and use the push/turn control
to set the monitor temperature to 200°C.
MIN
AUTO
MAX
˚C
4. Setting the vacuum
Rotate push/turn control clockwise until the vacuum indication is
flashing.
Hold down the SET key and use the push/turn control to set the required
vacuum of 50mb.
After releasing the SET key the oven briefly flashes the vacuum setpoint.
The display then shows the actual pressure and the control starts to
control to the selected vacuum of 50mb.
mB