23.2 Electronic temperature monitor

23.2.1 Overtemperature protection

MAX

Range: up to 10°C max above nominal
temperature (for nominal temperature see
label)

MIN

MAX

AUTO

˚C

Using the push/turn control
select the symbol

MAX

-Symbol anwählen.
Hold down the SET key and
set the protection temperature
using the push/turn control.

23.2.2 Undertemperature protection

MIN

Range: from 10°C below minimum
temperature of oven to 10°C above
nominal temperature of oven (for nominal
temperature see label).

The low alarm cannot be programmed
above the value set as high alarm.

Where no undertemperature protection is
required, this has to be set to the lowest
temperature.

MIN

MAX

AUTO

°C

Using the push/turn control
select the symbol

MIN

.

Hold down the SET key and
set the protection temperature
using the push/turn control.

Note:

The temperature monitor can be set independently of the operating mode.
During ramp operation the monitor temperature must always be set sufficiently far above the maximum
working temperature.